Accucold ARS49ML
Brukerhåndbok
Summit Appliance ARS49ML Pharma-Vac Performance Series Refrigerator
Model: ARS49ML
1. Introduksjon
This manual provides essential information for the safe and efficient operation, installation, and maintenance of your Summit Appliance ARS49ML Pharma-Vac Performance Series Refrigerator. Please read this manual thoroughly before operating the appliance and retain it for future reference.
The ARS49ML is a 49 cu.ft. upright pharmacy all-refrigerator designed for demanding life science, vaccine, laboratory, and medical cold storage applications, ensuring precise temperature control and stability.
2. Sikkerhetsinformasjon
ADVARSEL: For å redusere risikoen for brann, elektrisk støt eller skade når du bruker apparatet, følg disse grunnleggende forholdsreglene:
- Les alle instruksjonene før du bruker apparatet.
- Ikke oppbevar eksplosive stoffer som spraybokser med brennbart drivmiddel i dette apparatet.
- Dette apparatet må være riktig installert og plassert i samsvar med installasjonsinstruksjonene før det tas i bruk.
- Connect to a properly grounded outlet only. See 'Grounding Instructions' in the installation section.
- Koble aldri fra apparatet ved å trekke i strømledningen. Ta alltid godt tak i støpselet og trekk rett ut av stikkontakten.
- Koble fra apparatet før rengjøring eller reparasjoner.
- Ikke bruk mekaniske enheter eller andre midler for å akselerere avrimingsprosessen, annet enn de som er anbefalt av produsenten.
- Ikke skade kjølemiddelkretsen.
- Ikke bruk elektriske apparater inne i matoppbevaringsrommene på apparatet, med mindre de er av typen anbefalt av produsenten.
- Sørg for god ventilasjon rundt enheten.
3. Produktet er overview
The Summit Appliance ARS49ML features a robust stainless steel exterior and interior, designed for durability and easy cleaning. It includes a sophisticated digital temperature controller and optimized forced air cooling for superior performance.




4. Installasjon
4.1 Plassering
Choose a location with a strong, level floor capable of supporting the refrigerator's weight (410 lbs. empty, plus contents). Ensure adequate air circulation around the unit. Avoid direct sunlight or heat sources. Maintain at least 4 inches of clearance from walls at the back and sides for proper ventilation.
4.2 Utjevning
The refrigerator is equipped with wheels for easy movement. Once in its final position, ensure the unit is level to prevent door misalignment and ensure proper operation. Use a level and adjust the leveling feet (if present, or use shims under wheels if necessary) until the appliance is stable and level.
4.3 Elektrisk tilkobling
Before plugging in the appliance, ensure the electrical supply matches the specifications: 115V AC/60Hz, 7.5 Amps. The appliance must be properly grounded to reduce the risk of electrical shock. Do not use extension cords or ungrounded adapters.
4.4 Første oppstart
After installation, allow the refrigerator to stand upright for at least 2-4 hours before plugging it in. This allows the refrigerant to settle. Once plugged in, allow the unit to operate empty for several hours (e.g., 24 hours) to reach its stable operating temperature before loading products.
5. Drift
5.1 Digital Temperature Controller
The ARS49ML features an intelligent microprocessor digital temperature controller. The operating control range is adjustable from +2°C to +10°C (+35.6°F to +50°F).
- Temperaturvisning: The digital display shows the current internal temperature. It can be viewed in Celsius or Fahrenheit to the nearest degree.
- Min/maks temperatur: The controller tracks and displays the minimum and maximum temperatures recorded, updated in 15-minute increments. Consult the controller's specific instructions for accessing min/max readings.
- Temperaturprobe: A separate temperature probe provides continuous product readout for improved accuracy. An additional sensor reads interior air temperature for superior control.
5.2 Temperaturjustering
Refer to the specific instructions provided with the digital controller for detailed steps on how to adjust the set temperature. Generally, this involves pressing a 'SET' button and using arrow keys to increase or decrease the desired temperature.
5.3 Alarmer
The unit is equipped with audible and visual alarms for high/low temperature excursions. If an alarm sounds, check the temperature display and take appropriate action to correct the issue. Consult the controller's manual for alarm silencing and reset procedures.
5.4 Forced Air Cooling
The optimized forced air cooling system ensures excellent temperature stability and uniformity throughout the cabinet. This system also facilitates rapid temperature recovery after door openings and product loading, minimizing temperature fluctuations.

5.5 Lasting av produkter
When loading products, distribute them evenly on the wire shelves to allow for proper air circulation. Avoid blocking the air vents. The adjustable shelves allow for flexible storage configurations.


6. Vedlikehold
6.1 Rengjøring
Regular cleaning of your refrigerator will help maintain its appearance and efficiency. Always unplug the unit before cleaning.
- Interiør: Clean the interior surfaces and wire shelves with a mild detergent and warm water solution. Rinse thoroughly and dry with a soft cloth. Do not use abrasive cleaners or sharp objects.
- Ytre: Wipe down the stainless steel exterior with a soft cloth and a stainless steel cleaner or mild soap and water. Avoid harsh chemicals that can damage the finish.
- Kondensator spoler: Periodically clean the condenser coils (located at the bottom/rear of the unit) to ensure efficient operation. Use a vacuum cleaner or brush to remove dust and debris.
6.2 Tining
The ARS49ML features an automatic defrost system. This means that frost buildup on the evaporator coils is periodically melted and drained, eliminating the need for manual defrosting. Ensure the drain pan is clear and not obstructed.
7. Feilsøking
Før du kontakter kundeservice, vennligst sjekk følgende vanlige problemer:
| Problem | Mulig årsak | Løsning |
|---|---|---|
| Kjøleskap fungerer ikke | Not plugged in; Circuit breaker tripped/fuse blown; Power switch off | Ensure plug is secure; Reset breaker/replace fuse; Turn power switch on |
| Temperaturen er for varm | Door left open; Door gasket faulty; Excessive door openings; Unit overloaded; Poor ventilation; Set temperature too high | Close door fully; Check/replace gasket; Reduce door openings; Do not overload; Ensure proper clearance; Adjust set temperature lower |
| Overdreven støy | Unit not level; Objects vibrating against unit; Fan obstruction | Level the unit; Remove vibrating objects; Check for fan obstruction |
| Alarmen lyder | High/low temperature excursion; Door ajar | Check temperature and product integrity; Ensure door is fully closed; Consult controller manual for alarm reset |
8. Spesifikasjoner
| Trekk | Spesifikasjon |
|---|---|
| Modell | ARS49ML |
| Merke | Accucold (Summit Appliance) |
| Kapasitet | 49 kubikkmeter |
| Dimensjoner (H x B x D) | 83.75 cm x 55.25 cm x 31 cm |
| Dybde med håndtak | 32.75" |
| Dybde med dør på 90° | 56.5" |
| Innvendige mål (H x B x D) | 51.5 cm x 50.5 cm x 27 cm |
| Weight (Net/Shipping) | 410 lbs. / 430 lbs. |
| Avrimingstype | Automatisk |
| Dør/skapmateriale | Rustfritt stål |
| Electrical (Voltage/Frekvens/Amps) | 115V AC / 60Hz / 7.5A |
| Termostat type | Digital |
| Kjølemedietype | R290a |
| Shelf Type / Quantity | Wire / 8 (Adjustable) |
| Sikkerhetssertifiseringer | UL, ULC, ETL-S |
| Hjul | 4 |
9. Garanti og kundestøtte
For warranty information and customer support, please refer to the documentation provided with your purchase or visit the official Accucold/Summit Appliance website. For immediate assistance, you may contact the seller, VetSupplyOnline, through the Amazon platform. Please retain your proof of purchase for any warranty claims.